Product Details

DY-HT3 Rapid heating Infrared Thermometer Calibration Instrument/ Standard blackbody stove(300-1200c)

Brand Name DA YAO
Place of Origin China
Model Number DY-HT3
Temperature Range 300-1200c
Usage Industrial
Theory Temperature Controller
Accuracy ±0.1c

Product Features

DY-HT3 Infrared Thermometer Calibration Instrument

1、The definition of black body furnace:A sort of object that can absorb external radiation energy completely as well as emit its total energy.

2、Overview:the major technique target:emissivity ,diameter of its mouth,temperature homogeneity and uncertainty of radiation temperature。So as to ensure product quality, generally speaking,it is designed in the light of different temperature range. What we say often-- black body is commonly known as ‘Artificial black body. Its emissivity comes close to 1,but is not equal to 1.

The main function of black body radiation source is to to produce standard radiation in a certain temperature.Therefore,it is mainly used for calibrating various radiation thermometer on the part of temperature measurement. Such as optical pyrometer,infrared thermometer and thermal infrared imager.

In industry,it is mostly applied to the field of temperature calibration.Black body furnace is the most important product. As to the calibration and testing of radiation thermometer,usually methods of comparison were introduced. Namely, through radiation source with high stability (usually black body radiation source and associated equipment,) Make a comparison between the two different temperature values shown by calibrator and the radiation thermometer that is being calibrated. Then, to determine whether it is up to par or give the final results. Standard black body source,known as calibrated radiation source, is directed primarily towards calibrating radiation thermometer,infrared thermometer and temperature radiation sensor.Our company has the world’s most advanced technology with a complete product range. The black body-series range has a wide temperature range. It is characterized by its compactness and portability.It is fit for every level of calibration bureau to calibrate radiation thermometer not only in the laboratory but also in the field.

3product features

1. More portablelighter and handiermore accurate

2.Human-computer interfaceChinese displayoperation and touchelegant appearance

3.Instant temperature settingsimple operationfull automatic heating powerheating up rapidlytemperature control

4).On the basis of customer’s  request,you can obtain language interface from anywhere in the world. It is convenient for exporting。

4Product advantage:

1Novel design of exterioran integrative structure of furnace body and controlling temperature unit.

2Isothermal blackbody cavity has only one mouth.The radiation from its mouth is uniformly distributed in accordance with optical spectrum.The emissivity is more than 0.995.

3Heating element is isolated from shell materials.Use special software to do pressure limiting and heating.Mains transformer with high power is not needed.

4Automatically heating up,safe and reliable,excellent thermal stability.

5Convenient operation, small volume,light weight and easy to carry.On part of calibration, it can be used not only in the laboratory but also in the fields.

6Use single chip microcomputer to control and analyse data, simulation and multilevel digital filter, and modular.

7Measured value and set point can be displayed in two rows.High precision,multi-functional,great anti-interference performance.

5Technical index

Model name

DY-HT3 

Operating Temperature

300-1200℃

operation mode

Touchscreen operation

Temperature resolution

0.1

Regulation mode

Manual operation in terms of 12 different temperature ranges.Self-regulation of Power Consumption and temperature difference

Diameter

Φ40mmIf special dimensions were required,it can be manufactured for you.

     Emissivity

0.995

Temperature stability

Less than 0.1 and 0.1%t(t means temperature of radiation source

Temperature uniformity

Less than ±0.15and 0.15%t(t means temperature of radiation source

Uncertainty of radiation temperature

±0.3%tt means radiation temperature

 Power supply

220VAC

 Power consumption

2000W

Volumeheight×width×Length

390mm×290mm×390mm

Weight

10Kg

Note:This instrument can be custom-built as client’s request.

5、AgendumJJG 856-2015 verification regulation of  working radiation thermometers.

6、Operation interface:

    

If you have any questions,please feel free to contact me!

Tel: +86 186 5382 5886

Email: tadmjd3@gmail.com