Specifications
1.CE anaesthesia machine2. anaesthesia ventilator
3. integrated CO2 absorption system
Aeonmed is professional anesthesia machine, ventilator manufacturer, medical pump, operating light and table manufacturer since 2001, after more than 10 years development, Aeonmed has been grow up to OR&ICU operating expert.
With owning more than 800 staffs and covering 10,000 square meters manufacturing base, Aeonmed is moving toward to be most famous hospital medical solution expert to assist hospitals fine best solution.
Our main product lines:
- Anaesthesia machine/anesthesia machine
- Transport and ICU ventilator
- Infusion pump and Syringe pump
- Operating table and lamp/light
- Ceiling pandent
- Medical solution, Hospital information and area gas system
Aeonmed is welcome your connection and communication, we believe we are the best choice for your medical solution.
. Classical multifunctional anaesthesia machine
. Dynamic compensation system for tidal volume
. Volume & pressure ventilation, SIMV mode with flow trigger
. CE marked, meet clinical safety requirement of EU
. Advanced electromagnetic valve ensures accurate tidal volume and compensation
. Waveforms displayed include: P-T, F-T and V-T
. Solid coating and anti-dust smooth surface
Physical & Environmental Specification
Dimensions (WxHxD): 824x1382x680mm
Weight: 110kg (including two vaporizers)
Storage temperature: -22~+55
Working temperature: +10~+40
Storage humidity: ≤93%
Electrical Specification
Mains: 100-240VA C, 50-60Hz
Battery: 24VDC, minimum90minPower consumption: ≤50W
Pneumatic Specification
Gas supply: O2, N2O: 0.28-0.6Mpa
Flowmeters : O2: 0-1L/min; 1-10l/min
N2O:0-1L/min; 1-10L/min
Gas supply: O2 deficiency alarm
Hypoxic guard system
O2 flush: 35-75L/min
Working mode: Closed, Semi-closed, Semi-open
Driven mode: Pneumatically driven and electronically controlled
Operating mode: Man. / Vent.
Safety valve: ≤8kpa
Ventilator Specification
. Patient type: Adult, pediatric
. Ventilation modes: IPPV, PCV, SIMV(flow-trigger), Manual
. Setting: Touch key & navigator wheel knob
. Tidal volume: 0-1600ml
. Ventilation frequency: 4-100bmp
. I:E ratio: 2:1-1:8
. Pressure control: 5-60cmH2O
. Ppeep: 0-20cmH2O
. Inspiratory plateau: 0-50%Ti
. Trigger sensitivity: 2-30L/min
. High pressure: 21-80cmH2O
. Low pressure: 0-20cmH2O
. Monitoring: Display: 5.7TFT color screen
Pressure: Paw, Ppeak, Pplat, Pmean
Volume: MV, Vt
Breath rate
Gas: FiO2
Compliance
Graph display: P-T, F-T, V-T waveforms
. Alarm: Pressure: Paw, high/low limit
Volumes: MV high/low limit, Vt high limit
Low O2 supply pressure
Apnea
Self-testing failure
Power failure
Alarm silence: ≤120s